Defibrillator Specifications
9650-0912-01 Rev. U ZOLL R Series Operator’s Guide A–3
Display
Screen type High-resolution, liquid crystal display (LCD)
Screen size 6.5 inches (16.5 cm) diagonally
Display format Nonfade moving bar display.
Sweep speed 25 mm/s
Viewing time 5 seconds
4 seconds with certain monitoring parameter options
Messages ERASING REPORT, INSERT CARD, REPLACE BATTERY,
LOW BATTERY, PERFORM CPR, NOISY ECG, RETRY ANALYSIS,
CHECK PATIENT, ANALYSIS HALTED, PRESS ANALYZE,
NO SHOCK ADV., SHOCK ADVISED, PRESS CHARGE, SELECT PADS,
SELECT ECG LEADS, SELECT DEFIB MODE,
VF ALARMS OFF, REMOVE SYNC, CHECK PADS, ATTACH PADS,
POOR PAD CONTACT, DEFIB PAD SHORT, PADDLE FAULT,
ECG LEAD OFF, USE PADDLE DISCHG, CANNOT CHARGE, RELEASE
SHOCK, PRESS SHOCK, 30J TEST OK, TEST FAILED, PACER
DISABLED, DEFIB DISABLED, SET PACE MA,
CHECK RECORDER, ANAYLZING ECG, FULLY RELEASE
Electrodes
Hands-free therapy
electrodes
• OneStep electrodes
• Pro
-padz
• Stat
-padz
• Pediatric Pedi-padz
®
Defibrillator
Waveform Rectilinear Biphasic
Energy selection
(delivered to a 50
load)
• ADULT: 1, 2, 3, 4, 5, 6, 7 8, 9, 10, 15, 20, 30, 50, 75, 100, 120, 150,
200 joules (through front panel buttons or paddle buttons)
• PEDIATRIC: 1, 2, 3, 4, 5, 6, 7 8, 9, 10, 15, 20, 30, 50, 70, 85, 100, 120,
150, 200 joules (through OneStep pediatric electrodes only)
Charge Time
• Less than 7 seconds with a new, fully charged battery (first 15 charges
to 200 joules), with or without AC power at rated mains voltage.
• For the fifteenth discharge at maximum energy, the charge time is less
than 10 seconds, with or without AC power at rated mains voltage.
Depleted batteries result in a longer defibrillator charge time.
• Less than 15 seconds when operating without a battery, using AC
power alone at 90% of the rated mains voltage.
• Less than 25 seconds from the initial power on, with a new, fully
charged battery pack (depleted by up to fifteen 200 Joule discharges)
or when operating without a battery, using AC power alone at 90% or
100% of the rated mains voltage.
• Less than 30 seconds from initial power on and after rhythm analysis
(advisory mode) with a new, fully charged pack (depleted by up to
fifteen 200 J discharges) and using AC power at 90% of rated mains
voltage.
Patient Impedance
Range
15 - 300 ohms
Energy display Screen shows selected and delivered energy.